Installation of measuring dielectric loss Tangent 3m3

Specify the price
Availability: under the order   Condition: new




1.1. PURPOSE METER

1.1.1. Installation of measuring dielectric loss of liquid dielectrics & laquo; Tangent-3M-3 & raquo; is intended to measure the dielectric loss tangent of transformer oil and other liquid dielectrics conducted according to GOST 6581-75.

1.1.2. Device & laquo; TANGENT-3M-3 & raquo; . Made by meeting the requirements of GOST 22261 for electronic measurement instruments group 2

1.2 OPERATING CONDITIONS

1.2.1 & laquo; TANGENT-3M-3 & raquo; has the following normal conditions of use:

ambient temperature, & deg; C

20 & plusmn; 5;

relative humidity,%

80 at 25 & deg; C;

atmospheric pressure, kPa (mm Hg..)

84 & ndash; 106 (630 & ndash; 795);

mains frequency, Hz

50 & plusmn; 1

1.2.2 Device & laquo; TANGENT-3M-3 & raquo; designed for operation under the following operating conditions of use:

ambient temperature, & deg; C

from 10 to plus 35;

relative humidity,%

to 80 with 30 & deg; C;

atmospheric pressure, kPa (mm Hg...)

< p> 70 & ndash; 106,7 (537 & ndash; 800);

mains frequency, Hz

.

50 & plusmn; 1

1.2.3 Limit conveying conditions:

ambient temperature, & deg; C,

from minus 20 to plus 50;

relative humidity,% to

95 at 30 & deg; C;

< p> atmospheric pressure, kPa (mm Hg..)

70 & ndash; 106,7 (537 & ndash; 800);

tossing:

& nbsp;

number beats per minute

8 0 & ndash; 120;

maximum acceleration, m / s 2

30.

1.2.4 operation of the device can be made by one operator, who has studied this document, passed training in the practical work with the device, and has a group of safety in installations with a voltage greater than 1000, not less than the third

1.3 SPECIFICATIONS

1.3.1 Device & laquo;. TANGENT-3M-3 & raquo; provides in normal conditions of use measuring dielectric loss tangent (tg & delta;) and the capacity of samples of liquid dielectrics (Cx) in the ranges of the main measurement error, below:

- Measurement of tan range of the dielectric loss angle of 1. 10 -4 to 0,99;

- maximum permissible absolute basic error in the measurement of the dielectric loss tangent (1) + (2. 10-4 + 0,05 tg & delta;)%; < p> - capacitance measurement range at a test voltage of 1.5 kV to 2 from 5 pF to 100 pF;

- maximum permissible absolute error in the measurement of capacitance at frequency of 50Hz 1 pF Cx + 0.01;

- maximum permissible absolute error in the measurement of capacitance at frequency of 54Hz 1 pF Cx + 0.03;

-limit deviation of the measured liquid temperature from the set temperature, & deg; C 1;

- maximum permissible absolute basic error in the measurement of the dielectric loss tangent (2) + (2,5 10-4 + 0,07 tg &. delta;);

- maximum permissible relative error in the measurement of RMS voltage at a frequency of 50 Hz in the range of voltage measuring from 1 to 2 kV +3%;

- maximum permissible relative error measuring RMS voltage at a frequency of 54 Hz in the voltage measuring range of 1 to 2 kV +3%;

- absolute temperature limit of permissible error of measurement of the test liquid dielectric +1 & deg; C

. -time heating to 90 & deg; C, min. 80 & plusmn; 20

is the volume of the cell, cc 60 & plusmn; 2

-voltage supply single-phase alternating current, 220 & plusmn; 22

-Power consumption, KW, not more than 0,6

1 - capacitance measurement for the whole frequency range at a test voltage of 50 Hz;

2 - capacitance measurement for the whole frequency range for the test voltage 54Hz

1.3.2 limit of tolerable additional measurement error caused by changes in ambient temperature from normal (20 & plusmn; 5 & deg; C) to either within the working area, does not exceed 10% of the maximum permissible main absolute measurement error for every 10 degrees . temperature variation

1.3.2 & laquo;. TANGENT-3M-3 & raquo; provides the output of measurement results on the liquid crystal display (LCD) display unit.

1.3.3 Relationship between the high-voltage measuring unit and the display unit is carried on a radio channel in the frequency range 845-945 MHz. Stability of the radio at a distance 5-7 m level is provided in transmitter output power of 1 mW.

1.3.4 Image numbers and symbols on the display distinguishably at least 0.8 m at an external illuminance of 10 to 1,000 lux .

1.3.5 & laquo; TANGENT-3M-3 & raquo; under normal operating conditions and the application is ready to operate for no more than 2 minutes. after switching on.

1.3.6 Duration of continuous operation is determined by power consumption electrochemical batteries used in the display unit. (When the battery capacity of 1800 mA & bull; h operation time display unit is not less than 10 hours.)

1.3.7 MTBF for the operating conditions of application

is not less than 4000 h <. /p>

1.3.8 Average life meter & ndash; at least 8 years.

1.3.9 Overall dimensions and mass components of the meter shown in Table 1.

Table 1.

< tr>

Name part

designation

Max., Ea.

dimensions, mm

Weight kg

Note

1

unit main < / p>

& laquo; TAN-BO & raquo;

1

737x393x192 < /p>

20.5

 

2

display module

& laquo; TANGENT-3M-3-MI & raquo;

1

192x94x42

0.4

& nbsp;

3

charger < td>

9-12V

1

80x76x52

< p> 0.2

300-900 mA

4

cell measuring

& nbsp;

3

Ǿ105x40

0.4

& nbsp;

5

INSTALLATION MEASUREMENT OF DIELECTRIC LOSS dielectric liquid & laquo; TANGENT-3M-3 & raquo; MANUAL

3M3-00.00.00RE

1 < td>

& nbsp;

& nbsp;

& nbsp;

6

The set of wires and cables

KP-1

1

Ǿ200x40

0.3

& nbsp; & nbsp;

1.4. PACKAGE

Device & laquo; TANGENT-3M-3 & raquo; supplied completeness, shown in Table 1.

List of cables included in the set of connecting cables CP1 shown in Table 2.

Table 2.

Modification

Name of cables and wires

Number., pieces.

Note: < tr>

CP1

The wire connecting the high-voltage

1 < td>

Length 1.5m

network cable

1

1.8 m

The ground wire

1

Length 1.5m

& nbsp; < td>

& nbsp;

& nbsp;

& nbsp;

1.5 DEVICE

1.5.1 Device & laquo; TANGENT-3M-3 & raquo; It consists of the main unit & laquo; TAN-BO & raquo ;, hereinafter & laquo; TAN-BO & raquo ;, consisting of a heating cabinet or thermostat measuring unit with a control unit and display module & laquo; TANGENT-3M-3-MI & raquo ;. Communication between the measurement unit and the display module is carried out by radio. The main unit & laquo; TAN-BO & raquo; It heats three samples of dielectric liquid to a predetermined temperature, measuring the parameters of liquid dielectrics and transmitting data to the display module & laquo; TANGENT-3M-3-MI & raquo; . Via radio

1.5.2 schematic block diagram of the main & laquo; TAN-BO & raquo; is shown in Figure 1, and its appearance and design & ndash; Figures 2 and 3

1.5.3 Appearance display unit & laquo;. TANGENT-3M-3-MI & raquo; shown in Figure 4.

The display module in a plastic housing type KM103. On the front side are placed - the LCD screen and the controls and indicators described in the caption label. The side wall is jack adapter charger (9-12V, 300-900 mA).

On the back wall there is an access cover to the compartment supply. Power & laquo; TANGENT-3M-3-MI & raquo; It is made from four cylindrical nickel-metal hydride batteries of AA size 1.2 voltage and capacity of 1800-2500 mA & times;. h

1.5.4 measuring cell apparatus shown in Figure 5. The cell is made of stainless steel. As used insulators ring of fluoroplastic. The cell can be easily disassembled and meets the requirements of GOST 6581-75. Details cells made with sufficient accuracy and allow for multiple disassembly, cleaning and drying

1.6 Work product

1.6.1 & laquo;. TANGENT-3M-3 & raquo; It represents a set of electronic devices, measuring the following parameters simultaneously:

- load current (complex quantity);

- load voltage (scalar);

- load voltage (vector quantity);

- test voltage frequency;

- temperature of the dielectric liquid sample;

- air temperature in an oven;

- heating time, etc

From these data, the electronic device disposed in the main unit & laquo; TAN-BO & raquo ;, calculates the capacitance Cx of the test object.. and tangent tg & delta dielectric loss; and transmits the data by radio to the display unit, where they are displayed on the screen. Furthermore, the microprocessor of the main unit heats the dielectric fluid sample to a predetermined temperature and applying a high voltage to these samples

1.7 Labeling

1.7.1 On the rear side of the main unit & laquo;. Tan- BO & raquo; It has a label with markings: & laquo; TANGENT 3m-3-BO & raquo; .. №HHH, 2010, where XXX - serial number

1.7.2 The front panel display unit applied marking & laquo; TANGENT-3m-3-MI & raquo ;. The rear panel has a label with markings, the manufacturing serial number and similarly n 1.7.1

1.7.3 Commercial name appliance & laquo;.. Tangent 3M-3 & raquo; applied to the front of the main unit & laquo; TAN-BO & raquo; Fig.2. and Fig.3.

1.8 Packaging

1.8.1 Packing should be done only after full alignment device temperature and the ambient temperature of the room where the packaging is made. The room in which is made preservation and packaging, must be clean, relative humidity should not exceed 70% at an ambient temperature of 20 & plusmn; 5 & deg; C

1.8.2 Blocks device to be packed must not have. damage anticorrosive coating must be reliably protected from the external environment

1.8.3 Blocks device is placed in cases of a polyethylene film which is hermetically sealed with adhesive tape type & laquo;. scotch & raquo;.

1.8.4 On the crates applied handling marks & laquo; brittle. WARNING & raquo ;, & laquo; PROTECT FROM MOISTURE & raquo ;, & laquo; TOP & raquo;.