kilovoltmeter RD-30
|
1 DESCRIPTION AND OPERATION h3>
1.1 Product purpose p>
1.1.1 kilovoltmeters RD-30 (hereinafter - RD-30) is designed to measure the DC and AC sinusoidal voltage of 50 Hz . in the range from p>
1 and 30 kV p>
RD-30 provides measurement: p>
& bull; rms ac voltage frequency of 50 Hz; p>
& bull; the average value of the DC voltage; p>
1.1.2 The RD-30 can be used in high-voltage test laboratories of electrical and energy profile, as well as research institutions p>
1.1.3 Implementation of this item is. GOST 17512-82, climatic category UHL 4.2 according to GOST 15150-69 and group 2 of devices in accordance with GOST 22261-82. p>
1.2 p>
1.2.1 RD-30 operating conditions products intended for operating under the following conditions: p>
& bull; Improvement in the absence of shaking, vibration, corrosive vapors of liquids; p>
& bull; Ambient temperature? C - from 5 to 40S; p>
& bull; the relative humidity of ambient air,% - 80; p>
& bull; atmospheric pressure, kPa (mm Hg.). - 84 & ndash; 106,7 (630-800); p>
& bull; nonsinusoidality coefficient measured AC voltage should be no more than 8%; p>
& bull; coefficient measured DC voltage ripple should be no more than 1%; p>
& bull; the distance from the high-voltage unit RD-BW-30 to conductive objects, which are grounded or at a potential - not less than 0.5 m; p>
Documentation for download of the device not available