wattmeter & nbsp; D5064 & nbsp; for active power measurement in single-phase ac electrical circuits and & nbsp; DC. . The moving part of the instrument is attached on extensions Calm & ndash; liquid. span> p>
Maximum permissible wattmeter additional error caused by the deviation of the voltage at 20% of its nominal value at a constant value of the measured power is equal to 0.5% of the final value of the measuring range. p >
Maximum permissible wattmeter additional error caused by the frequency deviation from the upper limit of the normal range of frequencies to any frequency value in the operating frequency range is + 0.5% of the final value of the measurement range. p>
Maximum permissible additional wattmeters error caused by a change in temperature from normal ambient (20 + 2) C and any temperature within the operating range is + 0.5% of the final value of the measurement range for every 10 C temperature measurement. p>
Settling time wattmeter readings does not exceed 4 seconds at a temperature (20 + 2) p>
product Characteristics:. strong> p>
accuracy 0,5
Nominal voltage, V 30; 75; 150; 300; 450; 600
Rated current 0.5 A; 1
Normal frequency region of 45 Hz - 65
Working frequency region of 65 Hz - 500
Rated power factor of 1
Nominal parallel circuit current, mA 3
serial circuit power consumption in the normal region frequently in & middot; a 0.415