Meter power transformers parameters & nbsp; K505 & nbsp; measures current, voltage and power in single-phase and three-phase three-wire and four-wire AC circuits at the uniform and non-phase loads p>
Basic specifications K505 instrument:. strong> p>
Ammeter:
accuracy class: 0,5;
final measurement value ranges A: 0, five; one; 2.5; five; ten; fifteen; 50; 100; 150; 200; 300; . 600 p>
Voltmeter:
of the measuring range, B: 30, 75, 100, 300, 450, 600;
accuracy class:
1,0 (30);
0,5 (for the rest) p>
measurement range wattmeter:.
nominal current, A: 0.5-10; 15-600;
Rated voltage:
30; 75; 15-600 (the currents for 0.5-10 A);
30-600 (for the current 15-600 A) p>
class of accuracy built into instrumentation td>
0,5 td> tr>
accuracy class current transformer UTT 5M td>
0,2 td> tr>
Rated current, A td>
Measured capacity at nominal voltage, kV td>
Note: td> tr>
30 td>
75 td>
150 In td>
300 td> < td> 450 td>
600 td> tr>
0.5 td>
0.015 td>
0.0375 td>
0.075 td>
0.15 td>
0.225 td>
0.3 td>
Without current transformer UTT 5M td > tr>
1 td>
0,03 td>
0,075 td>
0,15 td>
0,3 td>
0,45 td>
0,6 td> tr>
2,5 td>
0,075 td> < td> 0,1875 td>
0,375 td>
0,75 td>
1,125 td>
1,5 td> tr> < tr>
5 td>
0,15 td>
0,375 td>
0,75 td>
1,5 td>
2,25 td>
3 td> tr>
10 td>
0,3 td>
0,75 td >
1,5 td>
3 td>
4,5 td>
6 td> tr>
15 td>
0,45 td>
1,125 td>
2,25 td >
4.5 td>
6.75 td>
9 td>
With current transformer UTT 5M td> tr>
< td> 50 td>
1.5 td>
3.75 td>
7.5 td>
15 td>
22 5 td>
30 td> tr>
100 td>
3 td>
7.5 td>
15 td>
30 td>
45 td>
60 td> tr>
150 td>
4.5 < / td>
11,25 td>
22,5 td>
45 td>
67,5 td>
90 td> < / tr>
200 td>
6 td>
15 td>
30 td>
60 td>
90 td>
120 td> tr>
300 td>
9 td>
22,5 td>
45 < / td>
90 td>
135 td>
180 td> tr>
600 td>
18 td>
45 td>
90 td>
180 td>
270 td>
360 td> tr>
Rated power factor wattmeter td>
1 td> tr>
Rated current, A td>
nominal resistance, Ohm td>
nominal inductance mH td> tr>
0,5 td>
10 td>
2,4 td> tr>
< td> 1 td>
3 td>
0.7 td> tr>
2.5 td>
0.6 td>
0, 14 td> tr>
5 td>
0,12 td>
0,04 td> tr>
10 td>
0,05 td>
0,02 td> tr>
Nominal frequency region td>
40-65 Hz < / td> tr>
Workspace frequency td>
at currents up to 10 A 65-400 Hz (for 75);
65-500 Hz ( As for 150-600) td> tr>
Rated current sets of parallel chains between the clamp and the neutral conductor terminal of any of the phases at the free position phase Sequence button (the phase indicator is not included) td>
10.5 mA td> tr>
Maximum permissible error of a set of devices in the normal range of 40 to 65 Hz frequency when measuring currents up to 10 a inclusive (without UTT-5M), stresses from 75 to 600 V and the capacity corresponding to the respective values of current and voltage, immediately after switching on and in continuous load is td>
& plusmn; 0.5% of the final value of the measurement range td> tr>
At 30 limit the maximum permissible basic error set devices in the frequency region equal to the normal td>
& plusmn; 1.0% of the final value of the measurement range td> tr>
Response time instrumentation kit td>
is not more than 4 s. Td> tr>
dimensions td>
no more 505h330h190 mm. td> tr>
set weight (with UTT-5M) td>
no more than 15 kg. td> tr>
The kit K505 td>
set measurement K505 & nbsp; & mdash; 1pc .;
current transformer UTT-5M & nbsp; & mdash; 1 pcs .;