1. PURPOSE

1.1. Apparatus high voltage test AV 50 / 70-3 (hereinafter & ndash; & laquo; apparatus & raquo;) for:

  • Test insulation safeners (insulating rods, rubber boots, gloves, power and so on. n) variable high voltage power frequency adjustable within 0 -.. 100kV
  • test power cables and insulation solid dielectrics rectified electrical voltage, as well as for testing of solid dielectrics sinusoidal electric voltage frequency of 50 Hz <. / li>
  • determine the breakdown voltage of liquid dielectrics (GOST 6581-75) & nbsp;

    1.2. The apparatus is designed for operation under a canopy or smokers at operating air temperatures from minus 10 & deg; C to plus 40 & deg; C, 80% relative humidity at a temperature of 20 & deg; C and atmospheric pressure 84,0 & ndash; 106.7 kPa (630 & ndash; 800 mm Hg...)

    1.3.. The apparatus can be used in two ways:

    • stationary,
    • laptop

      1.4.. To use this machine requires two operators who have permits to work on electrical installations with voltage over 1000 V

      & nbsp;.

      2. SPECIFICATIONS

      2.1. Power & ndash; single-phase AC mains voltage of 220 & plusmn; 20 V, frequency 50 & plusmn; 2Hz

      2.2. Power consumption, not more, VA 2000

      2.3. Number of high-voltage sources - 4

      2.4. Parameters of the sources listed in Table 2.1

      Table 2.1

      Option

      source Type

      BWI-50-1

      BVI-50-2

      BVI-20

      BVI-1

      The largest AC output voltage, kV

      50

      50

      20

      1

      The highest DC output voltage, kV

      70

      -

      -

      -

      The highest output current mA

      & nbsp;

      35

      35

      35

      50

      Overall dimensions

      270h295h660

      270h295h660

      270h295h660

      225h160h195

      Weight kg

      35

      35

      35

      3

      2.5. Test the bath parameters:

      • The number of test electrodes & ndash; 4;
      • dimensions (stand), mm 670h470h1120
      • dry weight, kg, not more than 10

        2.6. The parameters embedded kilovoltmeter:

        • limits measuring AC voltage, kV - 1, 20, 50, 100;
        • reduced relative measurement error,%, no more than - 3 < /li>

        2.7. Embedded current meter Parameters:

        • Measurement limit (AC and DC), mA, 25;
        • reduced relative measurement error,%, no more than 3 < /ul>

          2.8. a high-potential power meter parameters ITVR-4:

          • measurement limit (AC and DC), mA, 10;
          • reduced relative measurement error,%, no more than 3

            2.9. The area occupied by the apparatus (recommended), m2, 4

            2.10. equipment weight, kg, no more 100

            2.11. Life of the equipment device & ndash; 10 years