Spectrum analyzer S4-60
|
S4-60 spectrum analyzer for the study of the spectrum of periodic signals of arbitrary shape and modulation as well as the spectra of stationary noise.
spectrum analyzer S4-60 strong> for observation and measurement of the relative levels of the spectral components of the continuous periodic signals and . recurring radio pulses of oscillographic indicator for storage CRT p> S4-60 spectrum analyzer allows for: p> The device consists of three units: the ultra-high frequency transducers YA4S-59, 60-YA4S and oscillographic display YA40-0830 p> Features spectrum analyzer S4-60: h3>
Main specifications. device S4-60: h3>
instrument frequency range from 0.01 GHz to 39,6
breakdown subbands following: 0.01-1.5; 1,45-3,45; 3,45-7,45; 7,45-12,0; 12,0-19,45;
19,45-39,6
limits of acceptable basic error measurement input sinusoidal signal frequency MHz ± (10-2 f +10) in the range from 0.01 to 1 ,
± 5GHz (10-2 + f 2) with integrated calibrator
10-2 ± f in the range of 1.45 to 39,6GGts
limits of acceptable frequency measurement error under the operating conditions MHz ± (1,5h10-2 f +10) in the range from 0.01 to 1.5 GHz
± (1,5h10-2 f +2) calibrator with built 1,5h10-2
± f in the range of 1.45 to 39,6GGts,
where f-frequency sinusoidal input signal
spans Nominal values are set in the range from 50 kHz to 1000MHz frequency range from 0.01 to 1.5 GHz and from 50 kHz to 2000 MHz in the frequency range from 1.45 GHz to 39.6 discrete increments 1,2,5
nominal values of maximum frequency bands automatic adjustment 1500 MHz frequency range from 0.01 GHz to 1.5
Nh2000 in the range of 1.45 to 39.6 GHz,
where N - number of the first working harmonic oscillator
Stripes Propus Kania level - 3dB should kHz 1,3,10,30,100,300
instability frequency oscillators for 10 minutes after a setup time in normal conditions no more 500 kHz, when included phase-locked loop system;
Nx 1h103 off when the PLL frequency deviation
parasitic device heterodyne mode in REVIEW, no more than 350 Hz, when the PLL is enabled;
not more 15h103 off when the PLL
limits of acceptable basic error relative attenuation step attenuator NOMINAL LEVEL,% ± 4 in normal conditions
± 6 in operating conditions, at the position of the attenuator from 2.5 mV to 8 mV;
limits of acceptable soft error relative attenuation of the attenuator nominal level,% ± 12
limits of acceptable basic error indicator logarithmic amplitude scale ranging from 0 to - 70 dB, dB ± 2 - in normal conditions
± 3 - in the working conditions of Vp
unevenness udno-frequency characteristics of the device for input of the mixer, not more than 10 dB
average noise level in the pass band of 1 kHz at the input of the mixer, not more dBm - 100 in the range of 0.01 - 1.5 - 3.45 GHz < br> - no more than 95 in the range of 3,45-7,45 GHz
not more so - in the range of 90 GHz 7,45-12,0
not more so - in the range of 85 GHz 12,0-19,45
not more so - in the range of 70 GHz 19,45-39,6
by input selector, dBm - 85 in the range of 1,8-3,45 GHz
- 80 in the range of 3,45-7 , 45 GHz
- 75 GHz range 7,45-12,0
relative interference level at the input of the mixer when exposed to the input of two sinusoidal signals of equal amplitude with a level of 0.1 mW and the value of mismatch between them of 2 MHz dB is not more than - 60 in the range of 0.01 to 1.5 GHz
not more than - 50 in the range of 1.45 to 3.45 GHz
relative interference level at the input of the preselector when subjected to two input signals of equal amplitude with a level of 0.1 mW and the value of mismatch between them 100 MHz, db not more than - 60 in the range of 1.8 to 12 GHz
interference level on the image channel at the input selector in the range from 1.8 to 12 GHz, db not more than - 60
wave device resistance in the frequency range 0.01 to 12 GHz, 50 Ohm < br> The device has three types of internal scan sweep
with different kinds of start - single Sequence;
- Automatic startup;
- Automatic startup internal synchronization with the mains frequency;
- launching of the envelope pulse signal to be analyzed scale of not less than 3 grid cells
CRT device provides specifications after time samoprogreva, not less than 1 hour
AC power line voltage (220 ± 22)
rate (50 ± 0,5) Hz < br> The power consumed from the network, VA 400
dimensions (each block), mm 488 x 173 x 580
Weight (each of the blocks), 30 kg
Components analyzer transducer YA4S-59 < br> converter YA4S 60
indicator YA40-0830 with block YA4S-54 p> Replaces: S4-25, S4-27Vozmozhnaya substitution: S4-77, S4-85